Calibrate
CalibratePrint and measure

Enter your readings

The Measure step takes the L*a*b* readings of your printed chart, ink by ink. Type them, paste them or import a measurement file, then check them on screen.

note
Before you start

You need a preset of your own and its printed chart. Built-in and locked presets are read-only.

Type your readings

01.

Open Measure

In Calibrate, click Measure in the top bar, or press 2.

02.

Choose an ink

In the Inks list on the left, click the ink you read. Each row shows how many of its 21 patches are read, for example 0/21. ↑ and ↓ move to the next ink.

03.

Type the paper patch

In Measure LAB on the right, click the first field of the 0% row, the paper. Type L*, press Enter, type a*, press Enter, type b*, press Enter. The reading is set and the next patch is ready.

04.

Carry on to 100%

Repeat for each patch. A row you leave untouched is skipped. To keep a shown value as a reading, click its check mark (Set value).

05.

Save

Click Save at the right of the top bar, or press Ctrl + S. Rows still marked Pending are set too, and Saved appears.

tip
Serigraph presets

Without films, a screen prints the solid or nothing. Read the 100% patch: the others stay Simulated. To read every tone, print the chart in films.

Read the patch tags

  • Measured: a reading you set.
  • Pending: typed or pasted, not set yet. Enter, Set value or Save sets it; Auto Save does not.
  • Interpolated: filled in between readings while Auto interpolation is on.
  • Linearized: written by Linearize.
  • Simulated: no reading yet.
  • A warning icon next to a tag marks a reading that looks out of place: hover it, then read that patch again.

Paste or import readings

  • Paste: copy the L*, a* and b* columns from a spreadsheet and paste them into any row. They fill that patch and the ones below, marked Pending. Check them, then press Enter or Save.
  • Import: click Import under the list and choose the file exported by your measuring software (CGATS, CSV or TXT). A file with one ink's readings goes to the ink selected in the list.
  • If several charts of this preset fit the file, Which chart did you measure? opens. Click the chart you read, then Import: its ink order tells which ink each row belongs to.
  • LAB imported gives the count. LAB imported — check it also lists what to look at, such as patches still unmeasured.

Check the result

  • Under the grid, switch views: Swatches, Transfer, Tone gain, All transfer, All tone gain. Press V for the next one.
  • In Preview on the left, Before shows the ink without the preset and After with your readings.
  • In Transfer or Tone gain, drag a point up or down to correct it: its L*a*b* is rewritten. The 0% and 100% points stay fixed.
  • Reset clears every reading of the ink. Linearize puts its curve on the diagonal. Undo either with Ctrl + Z.

Measure overprint opacity

01.

Enter the solid and the paper

Above Measure LAB, Overprint opacity takes readings from the Overprints chart (Metallic coverage for a metallic ink). Type the ink's solid on its own, then the Paper row, and press Enter.

02.

Add the overprints

Click Add an overprint, choose the ink printed underneath, then type the reading. Used by the Editor switches from Estimated to Measured. Use the estimate removes these readings.

If something goes wrong

  • Readings not set opens when you change ink or preset with Pending rows. Set & switch keeps them, Discard drops them, Cancel stays.
  • A row ending with “— not set” was refused, for example “a* is not a number — not set”. Correct the highlighted field.
  • LAB import failed · “This file holds spectral data only”: export L*a*b* (or XYZ) from your measuring software.
  • LAB import failed · “Could not match any of the … records to your inks”: check that the preset and the selected ink match the chart you read.
  • Paste not read: paste only the L*, a* and b* columns, or include their header row.
READY?

Build with Shadrum.

Start now Download